Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic
12.06 CS PS SP
General Startup 27
Page 27 of 66
Medical Solutions
Functional test of C-arm lift movement and parking brakes 2.3
Fig. 5: Function test
• Switch on the main system (C-arm).
• To move the C-arm downward, press the downward button.
• The lifting column moves to position 1 (1/Fig.5/p.27) and remains there automati-
cally. A stop signal (3 beeps) sounds simultaneously.
• To lower the lifting column further, press the downward button once again.
• The lifting column can then be moved by a further 7cm to the lowest point, position 2
(2/Fig.5/p.27). In this area a signal (3 beeps) sounds for safety reasons each time the
downward button is pressed.
• To move the C-arm upward, press the upward button. No signal sounds in this case.
• When both switches are pressed, the lifting motor moves in the initially selected direc-
tion.
• Loosen the different parking brakes and apply them again. Secure blocking of the rele-
vant movement must be assured.
• Document the function of the C-arm movements, brakes and warning signals in the test
protocol, "C-arm movement and emergency stop" .