Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic
12.06 CS PS SP
Specific System Settings 43
Page 43 of 66
Medical Solutions
Click <Go> and wait until the message “Ready” appears in the footer.
Fig. 21: Backup & Restore_Command_Backup_SW-Settings02
ASPIA settings
Select "ASPIA settings" under Packages.
Click <Go>
- Wait until "Ready” is displayed in the footer.
Exam set
Select "ExamSet" under Packages.
Click <Go>
- Wait until "Ready” is displayed in the footer.
Security settings
For systems with HIPAA configuration
Select "Security-Settings" under Packages.
Click <Go>
- Wait until "Ready” is displayed in the footer.
Main system
Select "MainSystem" under Packages.
- Wait until "Ready” is displayed in the footer.
Exiting backup
Close the window via <Home>.
Remove the backup CD-ROM from the CD-ROM drive.