SITRANS F M
8. Commissioning
8.6.1
Settings available
Parameter Factory settings Settings available
Password
Default value 1000
Password 1000 1000 ... 9999
Basic settings
Flow direction Positive Positive, negative
Q
max.
Dim. dependent Dim. dependent
- Volume units Dim. dependent m
3
, ml, l, kl, hl, Ml, ft
3
, in
3
, USG, USkG, USMG, UKG,
UKMG, USBBL
- Time units Dim. dependent Sec., min., hour, day
Tot al ize r 1 Fo rw ard Fo rw ard, reve r s e, net
- Totalizer 1 units Dim. dependent m
3
, ml, l, kl, hl, Ml, ft
3
, in
3
, USG, USkG, USMG, UKG,
UKMG, USBBL
Tota li zer 2 Rev e r s e Fo rw a r d, reve rs e, n et
- Totalizer 2 units Dim. dependent m
3
, ml, l, kl, hl, Ml, ft
3
, in
3
, USG, USkG, USMG, UKG,
UKMG, USBBL
Low flow cut-off 1.5 % 0 ... 9.9 %
Empty pipe Off Off, on
Error level Warning Fatal, permanent, warning
Output
Current output Off On/off, uni-/bidirectional, 0/4 - 20 mA
- Time constant 5 s 0.1 ... 30 s
Digital output Pulse Error, direction/limit, batch
1)
, frequency, pulse, error no.,
off
Relay output Error Error, direction/limit, cleaning, error no., off
Direction/limit switch Off 1 set point/2 set points, 100 ... +100%
- Hysteresis 5% 0.0 ... 100%
Batch
1)
Off
- Batch quantity 0 Dim. dependent
- Batch compensation 0 100 ... +100 m
3
- Batch counter Down Up/down
- Time constant 0.1 s 0.1 ... 30 s
Frequency Off 500 Hz, 1 kHz, 5 kHz, 10 kHz
- Time constant 5 s 0.1 ... 30 s
Pulse On
- Pulse polarity Positive Positive/negative
- Pulse width 66 ms 64 µs, 130 µs, 260 µs, 510 µs, 1.0 ms, 2.0 ms,
4.1 ms, 8.2 ms, 16 ms, 33 ms, 66 ms, 130 ms,
260 ms, 520 ms, 1.0 s, 2.1 s, 4.2 s.
- Volume/pulse Dim. dependent Dim. dependent
-Time constant 0.1 s 0.1 ... 30 s
Electrode cleaning Off Off/cleaning
- Cleaning cycle time 24 h 1 ... 240 h
External input
External input Off Batch, reset totalizer, freeze output, forced output, off
- Batch Start, hold/continue, stop, Q
max.
2
Sensor characteristics
Correction factor 1 0.85 ... 2.00
Language English English, German, French, Danish, Swedish, Finnish,
Spanish, Russian, Italian, Portuguese and Polish
Operator menu
Primary field Flow rate Flow rate, Totalizer 1, Totalizer 2
Title/subtitle line Flow rate Flow rate, Flow rate %, Q
max.
, Totalizer 1, Totalizer 2,
Tot al izer 1 r es et, Tot alize r 2 res et, Batc h s ta r t/pa us ed/
stop, Batch cycle counter, Batch cycle counter reset,
Sensor size, Sensor type, Error pending, Status log, Tag No.
1)
Batch is available on MAG 6000 only
The transmitter is delivered with factory settings ready to measure the actual flow.