5 Miscellaneous
5.1 Checks
PMF Cassette locking
• Check that the camera used is properly locked to the object tables.
PMA UI’s and SPEED Infos
• Check whether all relevant UI’s and SPEED Infos have been completed.
• Check that the customer performs backup regularly.
5.2 Final tests
SIE Protective conductor test
• Perform the protective conductor test according to ARTD-002.731.17...
PMP Cleaning / Damaged paint
• Clean the unit using the materials recommended in
Instructions for use for MAMMOMAT
1000/3000 Nova
(chapter
Cleaning, Disinfection
) and repair any damaged paint.
• System ON.
QSQ Final test exposure
Take one final test exposure.