Miscellaneous 5 - 1
Siemens AG Register 8 SPB7-230.101.03 Page 1 of 2 Opdimar
Medical Solutions Rev. 04 07.03 CS PS 24 Maintenance Instructions
5 Miscellaneous
5.1 Checks
PMF Cassette locking
Check that the camera used is properly locked to the object tables.
PMA UI’s and SPEED Infos
Check whether all relevant UI’s and SPEED Infos have been completed.
Check that the customer performs backup regularly.
5.2 Final tests
SIE Protective conductor test
Perform the protective conductor test according to ARTD-002.731.17...
PMP Cleaning / Damaged paint
Clean the unit using the materials recommended in
Instructions for use for MAMMOMAT
1000/3000 Nova
(chapter
Cleaning, Disinfection
) and repair any damaged paint.
System ON.
QSQ Final test exposure
Take one final test exposure.