General information 1 - 1
Siemens AG Register 8 SPB7-230.101.03 Page 1 of 2 Opdimar
Medical Solutions Rev. 04 07.03 CS PS 24 Maintenance Instructions
1 General information
1.1 Training
• Training workshop or at least 1 installation
1.2 Required documents
1.3 Required tools, measurement and auxiliary devices
• Protective ground wire tester (44 15 899 RV090)
• Stereo calibration phantom (included in the Opdimar delivery) (part No. 64 30 701)
• Acrylic plastic (PMMA = Poly-methyl-met-acrylate) plates:
3 plates of AEC calibration plexiglas 150 mm x 150 mm x19 mm (part No. 65 61 232) and
one plate calibration plexiglas, 150 mm x 150 mm x 9,7 mm (part No. 65 61 224)
• Grid object table
• Mammography bar pattern (5-20 lp/mm),
for instance Nuclear Associates Resolution Test Pattern (07-555), http://www.nucl.com
• Cleaning agents* specified in
Instructions for Use MAMMOMAT 1000/3000 Nova
,
chapter
Cleaning, Disinfection
• Standard service tools*
• Instructions for Use MAMMOMAT 1000/3000 Nova SPB7-230.201...
• Supplement to the Instructions for Use MAMMOMAT 3000-
Opdimar
(Included in the Opdimar delivery)
SPB7-230.203.13...
• MAMMOMAT 1000/ 3000/ 3000 Nova - Opdimar
Service Instructions
(Included in the Opdimar delivery)
SPB7-230.061.11 ...
All tools, measurement and auxiliary devices listed, with the
exception of those marked with " * ", are specified in the STC
(Service Tools Catalogue).
NOTICE