Procedure - Potentiometer R863 3 - 3
Siemens AG SPB7-230.841.03 Page 3 of 6 M300/1000/3000/3000 Nova
Medical Solutions Rev. 02 04.04 CS PS 24 Service
Attach and solder the wires onto the contact lugs of the new potentiometer. Place it into
its metal enclosed observing the position of the little pin (1/Fig. 3) which should prevent
its turning; apply the washer and the mounting nut and tighten it.
Push the carrier bracket with the ball bearing onto the potentiometer shaft, place the
assembly onto the upper slab, and tighten the two hex socket head screws.
Re-attach the upper cable clamp, if necessary; the wires should not touch moving parts
(the toothed belt).
Drive the slide manually down to the lower limit.
Fig. 4 Fig. 5
Fig. 6 Fig. 7