Siemens-Elema AB SRX-MK RX B7-230.092.01.01.02 MAMMOMAT 3000
Solna, Sweden REV.01 10.95 Collimator holder
Modification of collimator holder 2-3
Installation of measuring tools
1 Put the left/right measuring tool (1) on left/right side of the collimator holder (2) and
secure at the upper edge by means of the wedge (3).
2 Put the clamp (4) with screw (5) over the collimator holder and measuring tool, lock
with the screw against the measuring tool.
3 Measure free space (Y-axis) on both sides between measuring tool and the side wall
of collimator cover (6).
4 Measure free space (X-axis) on both sides between the collimator holder and tube
plate.
Figure 3 Installation of measuring tools
NOTE!
Use the table over registration of radiation field to record measured values.