Siemens-Elema AB SRX-MK RX B7-230.092.01.01.02 MAMMOMAT 3000
Solna, Sweden REV.01 10.95 Collimator holder
2-4 Modification of collimator holder
Work-up of holes
1 Remove collimator protective cover by unscrewing four Allen screws.
Figure 4 Removal of collimator protective cover
2 Disconnect cables X866, X867 and X890 from the collimator and cut off the cable
ties. Notice the connector and cable-tie positions to ensure the same positioning after
work-up.
3 Unscrew the four Allen screws holding the collimator with holder.
Figure 5 Removal of collimator holder
4 Fold down the collimator with holder and secure with a cable tie to prevent it from
hanging in the cables.
5 Cover the collimator with a plastic bag. Protect other parts of the Mammomat with a
plastic cover.