Notes on the error messages
-nconErr
It is also permissible to terminate the wiring with data sinks identified by ncon. However , it is
advisable to add the missing connections because the desired functions cannot run with un-
defined inputs.
If the configuring preselection level is exited by the Exit key (tA1), the flashing error mes-
sage ncon Err appears if data sinks (inputs) are still marked ncon. The configuring preselec-
tion level is not exited, the error should be corrected.
Corrections:
The error is acknowledged by pressing the Enter key (tA4). It returns to the configuring
mode FCon to the first data sink marked ncon, the error can be corrected.
Cancel:
If you want to cancel the connection prematurely, press the Exit key (tA1) again after the
error message so that the online mode is switched to. The previous switchings are then
saved in a non--volatile memory.
---PoSErr
Ending positioning with non--positioned (but defined) functions is allowed. If the configuring
preselection level is to be exited with the Exit key, the flashing error message --Pos Err ap-
pears for non--positioned functions. The configuring preselection level is not exited, the error
canbecorrected.
The error message is acknowledged by pressing the Enter key. It jumps back to the config-
uring mode FPos to the first positioning number marked by nPos. The error can be corrected
or the online mode can be switched to by pressing the Exit key.
-nPoSErr
Ending positioning with a positioning row with nPos gaps is allowed.
If the configuring preselection level is to be exited with the Exit key and nPos gaps still exist,
the flashing error message nPos Err appears. The configuring preselection level is not ex-
ited, the error can be corrected. The error message is acknowledged by pressing the Enter
key. It jumps back to the configuring mode FPos to the first positioning number marked by
nPos. The error can be corrected or the online mode can be switched to by pressing the Exit
key.
Error messages for the display area of the displays dd1, dd2, dd3, dA1, dA2
oFL Exceeding the display range (19999 or 999) of the displays dd1, dd2 or dd3
--oFL, (oFL) Dropping below the display range (--1999 or --199) of the displays dd1, dd2 or
dd3
Flashing 1st or last LED of the analog display dA1, dA2: dropping below or exceeding the dis-
play range.