2 Installing a Video Printer1
General 1
Safety notice 1
Tests and adjustments performed with radiation on are identified with the radiation warn-
ing symbol . During these types of adjustments, radiation protection must be worn.
Protective conductor test 1
After completion of all service work and tests and reattaching all covers, perform the pro-
tective conductor test according to ARTD - 002.731.17....
The protective conductor resistance must not exceed 0.2 ohms.
Additional notice 1
These instructions apply to retrofitting a video printer on a
SIREMOBIL Iso-C only. 1
Parts required 1
Measurement and auxiliary tools required 1
Tools required 1
• Tool kit (refer to ARTD / part 3)
• 1 set of Allen keys
When performing service work and tests, adhere to the product-specific
safety information contained in the documents, as well as the general
safety information contained in the ARTD, part 2.
• Retrofit kit for installing video printer 31 00 497 G5437
• Service PC
• Service software (refer to Log book)
• Connection cable (from service PC to host) 99 00 440 RE999
• Protective conductor meter 44 15 899 RV 090