Installing a Video Printer 5
Siemens AG Register 3 SPR2-230.033.02 Page 5 of 6 SIREMOBIL Iso-C
Medical Engineering Rev. 02 10.99 TD SD 24
• Switch the SIREMOBIL Iso-C off and on again to activate the new configuration.
For Memoskop C-E (without keyboard)
• Switch on the system.
• Connect the service PC to the service interface of the SIREMOBIL Iso-C.
• Install and start-up the MEMOSKOP service program on the service PC.
• Program "Hardcopy release" on SIREMOBL Iso-C systems with one monitor.
• Program "Hardcopy release" on SIREMOBIL Iso-C systems with two monitors.
• If there is an area dose product measurement device present (Diamentor), activate
"Dose on hardcopy output" as well.
• Transfer the parameters to the MEMOSKOP.
• Exit the MEMOSKOP service program.
Function test 1
• Call up a fluoroscopic image on monitor A or monitor B.
• Release an exposure at the operating panel of the basic system and at the monitor
trolley keyboard.
The video printer should print an image in both cases.
Final steps 1
• Switch off the SIREMOBIL Iso-C.
• Test the protective conductor using the protective conductor meter 44 15 899 RV 090.
Refer to ARTD - 002.731.17...
Changes as compared to previous versions 1
• Paragraph on page 3 and 4 "Power cable" changed.
TD SD 24 / Arnold TD SD 24 / Groß SMS Iselin / O’Donnell