SITRANS P310 with HART
28 A5E35075329-01, 05/2015
3. With a
pressure transmitter below the differential pressure transducer
, open both blowout valves one after the other
until the air-free liquid escapes.
In the case of a
pressure transmitter above the differential pressure transducer
, open both vent valves one after the
other until the air-free liquid escapes.
4. Close both drain valves
or vent valves .
5. Open the differential pressure valve
and the vent valve on the positive side of the pressure transmitter slightly,
until fluid escapes without bubbles.
6. Close the vent valve.
7. Open the vent valve on the negative side of the pressure transmitter
slightly, until fluid escapes without bubbles.
8. Close the differential pressure valve
.
9. Open the differential pressure valve
until the liquid emerges and then close it.
10. Close the vent valve on the negative side of the pressure transmitter
.
11. Open the differential pressure valve
by rotating it in ½ turn.
12. Check and, if necessary, correct the zero point (4 mA) when the start-of-scale value is 0 kPa.
13. Close the stabilizing valve
.
14. Open the differential pressure valves (
and ) completely.
5.4.4
Commissioning with vapor
Pressure transmitter
Drain valves
Stabilizing valve
Condensate pots
,
Differential pressure valves
Differential pressure transducer/Orifice plate
Differential pressure lines
Insulation
Shut-off valves
Figure 5-2 Measuring steam
Requirement
All valves are closed.