4.1 Gas analyzers OXYMAT 6/OXYMAT 61/OXYMAT 7
NOTICE
Magnetism
Devices of type OXYMAT 6 or OXYMAT 61 or OXYMAT 7 emit magnetic stray elds. This is why
reason magnetically sensitive devices should not be installed in the immediate vicinity of these
devices.
• Keep a distance in such cases. This distance can be up to 50 cm depending on the sensitivity
of the aected device.
4.2 Laser gas analyzers SITRANS SL and LDS 6
CAUTION
Damage to the optics
If sensors are to be purged, they must not be mounted until the purging medium is available
and connected.
• Never switch o the purge function after purging, otherwise the wedge optics can easily be
damaged.
Assembly/installation
4.2 Laser gas analyzers SITRANS SL and LDS 6
Safety instructions
12 Compact Operating Instructions, 09/2018, A5E45779403002-01