6.2 Gas analyzers OXYMAT 6 and OXYMAT 7
CAUTION
Excessive sample gas pressure
Device damage
At the time of delivery, the pressure switch for reference gas monitoring is set to approx. 0.2
MPa (rel.). If the sample gas pressure has to be changed by more than ±150 hPa due to process-
specic requirements, the reference gas pressure also has to be changed by the same amount.
Only then is sucient monitoring functionality of the pressure switch ensured.
To reset the pressure switching point of the reference gas pressure switch
• Contact the Service.
• Have the pressure switching point of the reference gas pressure switch adjusted by a
Siemens technician or someone trained for this scenario.
6.3 ULTRAMAT 23 gas analyzer
WARNING
Danger of chemical burns
The O
2
S sensor contains acetic acid, which leads to burns on unprotected skin.
Therefore do not use any tools when replacing the sensor module which could damage the
sensor due to sharp edges or squeezing.
If contact with the acid occurs nevertheless, rinse the aected skin immediately with plenty of
water!
WARNING
Danger of chemical burns
The H
2
S sensor contains sulfuric acid, which leads to burns on unprotected skin.
Therefore do not use any tools when replacing the sensor module which could damage the
sensor due to sharp edges or squeezing.
If contact with the acid occurs nevertheless, rinse the aected skin immediately with plenty of
water!
Commissioning, operation, maintenance and servicing
6.3 ULTRAMAT 23 gas analyzer
Safety instructions
18 Compact Operating Instructions, 09/2018, A5E45779403002-01