4.1.3 SITRANS TF
4.1.3.1
Unsuitable cables, cable glands and/or plugs
WARNING
Unsuitable cables, cable glands and/or plugs
Risk of e
xplosion in hazardous areas.
Use only cable glands/plugs that comply with the requirements for the relevant type of
protection.
Tighten the cable glands in accordance with the torques speci󹪝ed in Technical data
(Page 79).
Close unused cable inlets for the electrical connections.
When replacing cable glands, only use cable glands of the same type.
After installation, check that the cables are seated 󹪝rmly.
NOTICE
Use of line and cable entries made of plastic in hazardous areas
De
vice damage caused by impact at temperatures below -20 °C.
Make sure that the line and cable entries are protected from impacts.
4.2 Installing SITRANS TH100/320/420 in a connection head
4.2.1
Installing SITRANS TH100/320/420 in the base of the connection head
Condition
The transmitter is only designed for installation in a type B connection head or larger.
Installing/mounting
4.2 Installing SITRANS TH100/320/420 in a connection head
SITRANS TH100/TH100Slim/TH320/TH420/TR320/TR420/TF320/TF420
Compact Operating Instructions, 07/2021, A5E41865021-AE 31