Instrument Measurement EC mAV 2
- 36 - © SIKA Ba_ECmAV2 03/2015
6 Instrument Measurement
CAUTION! Risk of injury and material damage!
Any voltage higher than 30 V will lead to possible personal injury and damage to the de-
vice.
Never apply more than 30 V between any two terminals, or between any terminal and
ground.
CAUTION! Material damage!
Exceeding the measuring range will cause possible damage to the device.
Do not apply a voltage or current exceeding the measuring range to the input termi-
nal.
CAUTION! Material damage!
Any connection with a device without cutting off its power supply will cause possible
damage to the device.
When connecting to the device, the power supply of the device under test should be
cut off.
6.1 Input Operation Procedure
Function operation % operation Display Measuring range
I = 0…20 mA DC
mA
mA %
00.000 mA
-0.25.00 mA%
-1.000…22.000 mA
-31.25…112.50 mA%
U = 0…28 VDC
0.000 V
-0.2000…28.000 V