4.5 Location of Acupuncture points with the Compact – Laser plus:
4.5.1 Preparation of the assembly
As well as the Silberbauer Compact - Laser mini (without power reducing tip!)
a Silberbauer Point Detector PS3 and a connection cable is required.
Put one plug of the connecting cable into the rear socket of the laser. The
other cable plug is required to be inserted into the golden socket of the Silberbauer
PS3 instead of the point-search-tip.
The point finder Silberbauer Punktsucher PS 3 must now be handed over into
the hand of the patient. The laser-tip can now serve now as a point finder.
4.5.2 Acupuncture point detection on the body
The Silberbauer Point Detector PS 3 indicates the electrical conductivity of the
skin both optically (with higher or lower frequency of the installed light-diode) and
acoustically (with a varying tone pitch). The latter can be heard by the doctor while
focussing his eyes on the acupuncture point.
Turn Rotary Switch 1 in position „Acup. Child“ or „Acup. Adult“. Set the laser
point close to the acupuncture point which you want to seek, slightly diagonally onto
the skins surface. The area assumed to be an acupuncture point is sought out; then
search the area, where you suppose the acupuncture point is, without lifting the tip
from the skin. The pressure of the tip should be light and consistent while
searching.
Continue the search in the direction of a higher tone or a quicker signal. The
tone pitch reaches its maximum light and maximum pitch or frequency at the centre
of the acupuncture point.
The push-button must be pressed on the laser above the acupuncture point
thus beginning the irradiation. The time of therapy programmed in the Silberbauer
Compact-Laser plus is optimised for this kind of application required in every model.