1 Intended Use
The Compact Laser plus is a Soft Laser (Low Level Laser) for radiation of
skin, mucous membrane and dental applications in human and veterinary medicine
(LLLT = Low Level Laser Therapy).
It is a “Medical Product“ in accordance with the EU Medical Product Law and
with the Essential Requirements 93/42 EWG and 2007/47 EC. This unit is also
approved by TGA.
The main applications are divided into the parts
Acceleration of wound healing and to avoid obstructions in healing, especially
for older patients, for diabetics and patients with wound healing
complications,
Reduction of pain e.g. after traumata and post surgeryand at chronic-
degenerative diseases,
Acupuncture instead of needles.
2 Safety precautions to be observed during the use
of therapy lasers
The appropriate legal security precautions are to be observed!
Direct irradiation of the opened eye should be avoided at all costs! Irreparable
damage can be expected!
even the closed eye should not be irradiated!
During radiation to the face use Soft Caps!
Laser therapy should be conducted by trained personnel only!
The use of the operating facilities or installation contingencies in any methods other
than mentioned in these user instructions can lead to dangerous irradiation!
CL plus- Laser must be used and stored within the following temperature range:
Case Temperature: during use 10 to 45
o
C (50 to 112
o
F),
Storage temp. 5 bis 50
o
C (40 to 122
o
F)!
Being the maximum temperature of the crystal, it is required!
Caution is recommended during its transport by car in the summer!
Permissible humidity: 30 till 95 %
rel
,
The humidity should not condense on the case/body (no dew-drops)!
The use of inflammable anaesthetic gases or oxidized gases like nitrogen (N
2
O) and
oxygen should be avoided. Some materials like cotton, that are saturated with oxygen,
could be set on fire at these high temperatures, created at the because of the laser’s
direction. Before the laser is put into operation, there should be a specific period for
evaporation of solvents in adhesives and inflammable solvents, which are used for
cleaning and disinfection. Attention should be drawn to the fact that the body’s gases
too can set on fire.
It is further recommended that:
Therapy duration for babies being treated at the cranial region should be short (approx. 50%)!