Compact - Laser Series CL plus Instructions for use
Energy: in Joules (J) = Watt-seconds (Ws)
Power: in milli-watt (mW) 1 mW = 0,001 W
Time: in seconds (s)
Unit area: in cm²
Irradiation-doses: in Joule / cm²
4.6.3 Calculation of the correct dose
Checking the area of a wound with the card:
Place the card close to the wound. Seek a circle which has approximately the
same area like the wound.
Close to this circle you will find a number which shows the size of this area in
cm² . Set the size switch at the laser in this position.
4.7 Checking the output power of the laser
The Silberbauer Compact - Laser plus has a mechanism for control of the laser
output power:
A yellow lamp which glows only if the laser works between 90% und 110% of ist
nominal output power, and a red one which will blink 2x per secound if the power is
too low, and which will glow permanently together with a 5 secound long warning