The output power should be: in position „Acup. Child“ 25% of the nominal output
power, in Position „Acup. Adult“ 50% and in all other positions 100% of the nominal
output power.
Range of tolerance of the measured power: Nominal Power +/- 5%
If the measured power is out of the mentioned range of tolerance, the laser must be
calibrated or repaired.
3. Checking the readability of all labels
All labels must be perfectly readible!
4. Checking of all accessories
Inspection of all accessories by visual inspection for breaks or wear should be
done. Also goggles for deep scratches which disturb the view.
9.2 Inspection Certificate
To confirm the accurate functioning of the instrument the owner will receive an
inspection certificate including its measured power. The results are given in the
instruction book.
10 Laser dispatch
The laser is delivered in its original packing, devoid of any defects by post as bulk goods.
Pull the battery out of the laser during transport!
Storage temperature must not exceed 50 °C (122 °F) !