Operating Manual TRACSCAN 2.0
V2.0, 2022-05 15 / 36
Test Program NFSI B101.3-2020
Parameter Name Default Value Can be altered
Distance 15 cm
Required Test Foot Type SBR, see section [4].
Table 6: Test Program NFSI B101.3-2020
Test Program NFSI B101.1-2020
Parameter Name Default Value Can be altered
Scale / Standard NFSI B101.1-2020
COF Type SCOF
Wet Measurement Yes
Ramp Level, Only
Residence Time 0.1 s
Number of Samples 5
Discard 1
st
Sample Yes
Lift / Lower Between Samples No
Required Test Foot Type SBR, see section [4].
Table 7: Test Program NFSI B101.1-2020
Test Program ANSI A326.3-2017
Parameter Name Default Value Can be altered
Scale / Standard ANSI A326.3-2017
COF Type DCOF
Wet Measurement Yes
Ramp Auto Detect
Residence Time 0.1 s
Distance 15 cm
Required Test Foot Type Any
Table 8: Test Program ANSI A326.3-2017
Test Program German Ramp
Parameter Name Default Value Can be altered
Scale / Standard German Ramp