Sirona Dental Systems GmbH 3 Electromagnetic compatibility
Installation Requirements C3
+
, C4
+
, C4
+
Cart
60 27 945 D3382
D3382.021.02.12.02
33
båÖäáëÜ
The UNIT is intended for operation in the electromagnetic
environment specified below.
The customer or user of the UNIT should make sure that it
is used in such an environment.
3.2 Electromagnetic emission
Emission measurement Conformity Electromagnetic environment guidelines
HF emission according to CISPR 11 Group 1 The UNIT uses HF energy only for its internal function. The
HF emission is therefore very low, and it is improbable that
nearby electronic devices might be disturbed.
HF emission according to CISPR 11 Class B The UNIT is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
dential purposes.
Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations / Flicker according
to IEC 61000-3-3
compliant