26 • ENGLISH
ELECTROMAGNETIC COMPATIBILITY
Table 3 —PROXIMITY FIELDS FROM RF WIRELSS
COMMUNICATIONS EQUIPMENT
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specied below.
The customer or the user of this monitor should assure that it is used in such an environment.
Test frequency Band Immunity test levels
(MHz) (MHz) Professional healthcare facility environment
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785