26 • ENGLISH
TROUBLESHOOTING
Problem Possible Cause Solution
Cu position was not correct or was
not tight enough.
Body posture was not correct during
measurement.
Speaking, excessive movement, angry,
excited or nervous anxiety during
measurement.
Batteries may be expired or installed
improperly.
Incorrect operation; possible strong
electromagnetic interference.
Display shows an
abnormal result.
Display is blank when
power is on.
No response when
POWER is pressed
or when new batteries
have been loaded.
Monitor keeps re-
inating.
Apply the cu properly and attempt a new
measurement.
Review the ‘Taking Your Blood Pressure Reading’
portion of this manual and attempt a new
measurement.
Wait a period of time and attempt a new
measurement after calm has returned; do not
speak or move during measurement.
Check the polarity of the batteries and reinstall if
necessary; replace batteries with new batteries.
Remove batteries and wait a minimum of ve
minutes; reinstall batteries and attempt a new
measurement.
The battery is low.
Replace the batteries.
Circuit is locked. Remove and reinsert the batteries;
attempt new measurement.