ERROR CODES
Pressure system is unstable before
measurement.
Failed to detect systolic/ diastolic pressure.
Pneumatic system leakage or cu is too loose/
too tight during ination.
The cu pressure is above 300mmHg.
More than 3 minutes with cu pressure above
15 mmHg.
EEPROM accessing error.
Device parameter checking error.
Pressure sensor parameter error.
E0
E5
E6
E7
E8
ER A
Re-apply the cu and attempt new
measurement.
Sit comfortable and remain still; ensure the cu
is at the same level as your heart.
Re-apply the cu and attempt new
measurement.
The measurement range is over 300 mmHg;
retake measurement and if error persists, consult
your physician.
Re-apply the cu and attempt to take a new
measurement. Sit comfortable and remain still;
attempt new measurement.
Code Meaning
Corrective Action
E1/ E2
E3/ E4