ENGLISH • 27
ERROR CODES
Pressure system is unstable before
measurement.
Failed to detect systolic/diastolic pressure.
Pneumatic system leakage or cuff is too
loose/ too tight during ination.
The cuff pressure is above 300mmHg.
More Than 3 minutes with cuff pressure
above 15 mmHg.
EEPROM accessing error.
Device Parameter checking error.
Pressure Sensor parameter error.
E0
E1/E2
E3/ E4
E5
E6
E7
E8
ER A
Re-apply the cuff and attempt new
measurement.
Re-apply the cuff and attempt new
measurement.
Wait 5 minutes. Sit comfortable and
remain still; attempt new measurement. If
the error persists, please contact Customer
Care.
Code Meaning
Corrective Action
TECHNICAL ALARM DESCRIPTION
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specied in part SPECIFICATIONS. In this case, you should consult a physician or
check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and
cannot be adjusted or inactivated. This alarm condition is assigned as low priority
according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.