TROUBLESHOOTING
1) System cannot achieve prime (water does not reach pump)
The particulate load in the water may be too high for the given filter pore size,
the max pressure is set too low, or the head pressure is too great. First, try low-
ering the elevation of the pump. Next, try increasing the pump pressure. If you
still cannot achieve prime, try a larger pore size filter.
2) The remote control display light is flickering or remote not communicating
with pump.
Change the remote control batteries. It’s advised to carry spare AA batteries
and a #1 Phillips screwdriver when sampling.
TIPS AND TRICKS FOR LONG TERM OPERATION
• Follow the pump storage procedure guidelines (do not store dry for pro-
longed periods).
• Use the caps provided for “IN” and “OUT” ports when system is not in use.
• Avoid getting water inside the remote control battery compartment.
• Do not twist the bipod legs (use the button on top to extend or retract).
For video tutorials and updates, please visit:
www.smith-root.com/support/tutorials
For additional assistance with your eDNA Sampler system, please contact Smith-
Root at 360-573-0202 or visit www.smith-root.com.