Ensure that the pump is stopped, the tubing is
disconnected from the patient, and the clamp
is open on the patient tubing.
1. Press Tasks use U and D to highlight the
Prime Tubing option, then press S.
2. When prompted, unlock the pump.
3. Press Prime to start priming.
4. Press Stop Priming when the air is removed, or
the delivery will stop automatically at 10 ml.
Prime the tubing
You can start and stop the pump, and deliver a PCA dose
(if one is programmed) without unlocking the pump.
For all other tasks, you must enter the security code.
When the security code is required,the pump displays
Enter Code to Continue.
1. Use U and D until the first digit of the
code is shown on the screen.
2. Press S to advance to the next digit.
3. Repeat with the second and third digits,
then press S or Accept Value to finish
entering the code.
Note: If the pump displays Use Key or Code to Continue, you can
unlock the keypad using the pump key, or enter the code, as above.
Unlocking the pump

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals