There may be an
internal device issue.
This icon is solid red on the Control Panel
and the device is not running.
Unplug your SoClean 3 from its
power source and contact SoClean
Customer Care.
SoClean 3 will
not operate a
Maintenance
Cycle or Fresh
Air Cycle
Your SoClean 3 is pumping air through
the system to remove ozone and allow
you to access your sleep equipment
mask and hose. Allow this process to
complete. You can either remove your
mask and hose for use or start a new
Maintenance Cycle when the green
Cycle Complete icon appears.
NOTE: Your sleep equipment mask
and hose may not have been fully
maintained before this issue occurred.
Wait a minimum of 30 minutes between
completion of the Cycle and use of the
sleep equipment mask and hose.
Your SoClean 3 was operating
a Maintenance Cycle and one
of the following occurred:
• Power to the SoClean 3
was momentarily lost
• Lid was forced opened
• The sleep equipment hose
was not detected after the
Maintenance Cycle started
• Internal Device Issue
The SoClean 3 is
making a noise
and these
indicators are
illuminated red on
the control panel
NOTE: These icons
may or may not be
illuminated on the
Control Panel.
None of the lights illuminate,
even when you press buttons
The SoClean 3 may not
be connected to power.
Check that the Power Supply is
plugged in correctly into the back of
the SoClean 3 and the AC wall outlet.
If the issue persists, contact SoClean
Customer Care.
SoClean 3 will
not operate a
Fresh Air Cycle
Fresh Air button
flashes when
pressed.
You have run 3 consecutive
Fresh Air Cycles.
Run a Maintenance Cycle.
POSSI BLE CAUSES SO LUTIONSPROBL EM YOU SE E
G E NE R AL TR O U B L E S H O OT I N G
G E NE R AL TR O U B L E S H O OT I N G
44 45