What to do if the front legs of my Spencer CROSS Medical Equipment don't lock when unloading?
- GggallegosAug 17, 2025
If the front legs of your Spencer Medical Equipment do not lock when unloading, it could be due to several reasons. The moving mechanisms might be damaged, in which case you should immediately remove the device from service and contact the service center. It's also possible that the loading floor height is not suited to the device, so adjust it to meet the manual's requirements. If adjustments aren't possible, remove the device from service and contact the service center. Another cause could be that the magnet controlling the release has moved; manually unlock the stretcher, check the magnet's position, and restore it. If the problem continues, remove the device from service and contact the service center.