EasyManua.ls Logo

Spirax Sarco EP500 - Regular Maintenance; Corrective Maintenance

Spirax Sarco EP500
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IM-P343-44 CTLS Issue 522
EP500 Standard Electropneumatic Positioner
5.1 Regular maintenance
1. Drain any build-up within the air supplylter set as impurities such as oil, water and dirt will
cause inconsistent operation.
2. Ensure air supply is at the correct pressure (see Section 3.3.2 and refer to the actuator TI).
3. Make visual checks to ensure that the valve is operating correctly.
5.2 Corrective maintenance
5.2.1 Removal and cleaning of sensitivity adjuster (see Figure 22):
-
Loosen and remove the lock.
-
Note setting then remove the sensitivity screw.
-
Wash the adjuster with solvent, checking the condition of the cone and ensuring that the
hole is clear.
-
Dry with clean compressed air to ensure that there is no remaining contaminants.
-
Ret the the cleaned adjuster and unscrew it ¼ turn from its stop.
Fig. 22
-
Ret the lock to touch the adjuster top and secure it with the lock-nut.
-
Set the sensitivity see Step 2, Section 4, Commissioning.
-
Reset zero and span if required.
Ensure that the side hole is clear
5. Maintenance

Related product manuals