Questions and Answers

  • D
    Daniel SmithAug 17, 2025
    Why are my specimens uneven with the Struers Control Panel?
    • L
      laurenhowellAug 17, 2025
      Uneven specimens can occur if the specimens are wider than the radius of the preparation disc. Try using smaller specimens. Also, check if the specimens are passing over the center of the disc; if so, reposition the horizontal position of the control panel.
  • J
    Jordan GarciaAug 20, 2025
    What should I do if the column starts to turn on my Struers LaboForce-100?
    • J
      Jose JacksonAug 20, 2025
      If the column starts to turn, it's likely due to loose column screws. Tighten the screws immediately.
  • B
    Bonnie BoothAug 22, 2025
    How to fix a vibrating specimen holder plate on a Struers Control Panel?
    • K
      Kristin GriffinAug 23, 2025
      A vibrating specimen holder plate can be caused by either an unbalanced plate or loose screws. You can try the following: * Replace the specimen holder plate if it's unbalanced. * Tighten the specimen holder plate screws if they are loose.
  • T
    Tina GeorgeAug 26, 2025
    What to do if the surfaces are being damaged in Struers Laboratory Equipment?
    • A
      Angela McclureAug 26, 2025
      If the surfaces are being damaged when using Struers Laboratory Equipment, it could be due to several reasons. The specimen may have burrs or sharp edges that need to be removed. Alternatively, the force applied to the specimen might be too high, or the specimen area might be too small. Try removing burrs and sharp edges, and lower the applied force to resolve this issue.
  • C
    caseycamposAug 28, 2025
    How to fix 'process cannot be started' error in Struers Laboratory Equipment?
    • M
      Molly AliAug 28, 2025
      If the process cannot be started on your Struers Laboratory Equipment, and you are seeing error messages, it might be because the flexible specimen holder is too close to the surface. To fix this, try changing the force, using higher specimens, or changing the configuration.
  • J
    Justin CisnerosAug 31, 2025
    What to do if pins are touching the surface on Struers Laboratory Equipment?
    • A
      angela99Aug 31, 2025
      If the pins on the flexible specimen holder are touching the preparation surface while using Struers Laboratory Equipment, it's likely that the force setting is too high, or the specimens are too low. You can decrease the force or increase the specimen height.
  • D
    Dawn ColeSep 2, 2025
    Why is the polishing cloth life so short on Struers Laboratory Equipment?
    • T
      Tricia AlvaradoSep 2, 2025
      If you are experiencing a short lifetime of the polishing cloth with your Struers Laboratory Equipment, it might be because the polishing cloth is too soft. Try using a harder polishing cloth.
  • A
    Amber TownsendSep 5, 2025
    What to do if specimens fall out of Struers Laboratory Equipment?
    • K
      knappjesseSep 5, 2025
      If the specimens fall out of the flexible specimen holder in your Struers Laboratory Equipment, it could be due to incorrect placement or insufficient force. Increase the force.
  • J
    Joseph GarnerSep 8, 2025
    Why do specimens develop a pencil shape in Struers Laboratory Equipment?
    • C
      Cassandra WilsonSep 8, 2025
      If the specimens develop a pencil shape when using Struers Laboratory Equipment, it could be due to several reasons: incorrect placement, a sample area that is too small, a specimen that is too high, or a specimen geometry that is round and solid. Turn the specimen and repeat the step.
  • Y
    Yvonne BrownSep 12, 2025
    What causes pins to stick in Struers Laboratory Equipment?
    • B
      Brittney SandersSep 12, 2025
      If the pins on the flexible specimen holder get stuck to the bottom plate of your Struers Laboratory Equipment, it's likely because the flexible specimen holder is dirty. Clean the flexible specimen holder.

Summary

About this manual

Safety Information

Intended Use

Defines the purpose and compatible equipment for LaboForce-100.

Safety Precautions and Messages

Lists general/specific precautions and explains hazard symbols like DANGER, WARNING, CAUTION.

Getting Started

Device Description and Overview

Overview of the LaboForce-100 unit, its components, and front view diagram.

LaboForce-100 Operation Basics

Basic operation steps like starting and stopping the machine.

Transport and Storage

Handling and Packaging Guidelines

Guidelines for preparing the unit for storage or shipping securely.

Installation Procedures

Unpacking and Packing List Verification

Steps for unpacking the unit and checking the box contents.

Initial Setup and Connections

Procedures for mounting, electrical, and compressed-air connections.

Specimen Holder and Plate Installation

Procedures for inserting and adjusting specimen holders and mover plates.

Specimen Mover Height Adjustment

Steps to adjust the height of the specimen mover plate.

Horizontal Position Adjustment

Procedure for adjusting the horizontal positioning of holders/plates.

Completing Installation Adjustments

Final steps to secure the specimen mover adjustment.

Flexible Specimen Holder Setup

Instructions for setting up the flexible specimen holder.

Flexible Holder Insertion and Height Adjustment

Procedures for inserting and adjusting the flexible holder's height.

Flexible Holder Horizontal Positioning

Procedure for horizontal adjustment of the flexible holder.

Accessory Integration

Using LaboDoser-100 and LaboDoser-10 with LaboForce-100.

Noise and Vibration Information

Reference to finding information on noise and vibration levels.

Operating the Device

Control Panel and Button Functions

Explanation of buttons, controls, and the touch pad on the panel.

Maintenance and Service

Routine Cleaning Schedules

Recommendations for daily, weekly, and mover head cleaning.

Monthly Maintenance Tasks

Procedure for emptying the water/oil filter.

Service, Repair, and Spare Parts

Guidelines for service checks, repair, and obtaining parts.

Service Information and Disposal

Accessing machine status info and equipment disposal guidelines.

Troubleshooting Guide

Error Identification and Causes

Table listing common errors, their causes, and recommended actions.

Error Message Classification and Handling

Classification of messages/errors and how to acknowledge them.

Technical Data

LaboForce-100 Specifications

Detailed specifications for motor, environment, power, air supply, and dimensions.

Diagrams and References

References to block and air diagrams for the LaboForce-100.

Legal and Regulatory Information

FCC Compliance and Notices

FCC notice and compliance information for the device.

Manufacturer Information

Manufacturer Responsibility and Disclaimers

Disclaimer regarding errors, changes, and manufacturer's responsibility.

Declaration of Incorporation of Partly Completed Machinery

Struers LaboForce-100 Specifications

General IconGeneral
BrandStruers
ModelLaboForce-100
CategoryControl Panel
LanguageEnglish