6. Move the adjustment ring up until the pin fits into the V-
shaped groove on the cabinet of the control panel.
7. Tighten the adjustment ring to fix it in this position.
8. Press the Turn/push knob knob to clear the error
message.
9. Press the Lower/Raise button to raise the specimen
mover head.
5.6.4 Adjusting the horizontal position of the specimen holder or specimen mover plate
With a specimen holder or specimen mover plate mounted:
1. Press the Lower/Raise button to lower the specimen
mover head.
2. Loosen the 2 fastening screws holding the column of the
control panel.
3. Replace the manual splash guard with a splash guard for
semi-automatic preparation or splash guard for Wet
grinding disc.
4. Moves the specimen mover head to the right.
With an MD-disc
1. Place the specimen mover plate in a position that allows
the specimen to run 3 – 4 mm over the edge of the
preparation disc.
A Disc edge
B Specimen mover plate
14 LaboForce-100
5 Installation