# Error Message Explanation Action
26 Information
Specimen mover plate not
lowered!
The pneumatic head with
specimen mover plate is not
down when the process is
started.
Appears if method is started
in single specimen mode
(SS) and mover plate is not
lowered.
Lower the pneumatic head.
Re-start the process.
9 Technical data
9.1 Technical data - LaboForce-100
Subject Specifications
Motor output 180 W
Rotational speed 50-150 rpm
Rotational direction Clockwise/Counter-clockwise
Force, individual
specimens
10-50 N
Force, specimen holder 30-300 N
Safety standards See the Declaration of Conformity
REACH For information about REACH, contact your local Struers office.
Noise level See the noise level stated for: LaboPol
Power supply LaboForce-100 is connected directly to LaboPol
Software and electronics Controls Touch pad, Turn/push knob knob
Memory FLASH-ROM/RAM/NV-RAM
LC display TFT-colour 320x240 dots with LED
back light
Operating environment Surrounding
temperature
5 – 40°C/41 – 104°F
Humidity < 85 % RH non-condensing
Storage and transport
conditions
Surrounding
temperature
0 – 60°C/32 – 140°F
Humidity < 85 % RH non-condensing