# Error Message Explanation Action
48 Error
Frequency inverter error!
The disc motor is
overloaded.
Fault code: 0.0
Thermal level: 0%
Wait for the disc motor to
cool.
Reduce the force and
continue the preparation
process.
49 Error
The disc motor is
overheated!
Please wait some minutes
and reduce the load.
Fault code: 0.0
Wait for the disc motor to
cool.
Reduce the force and
continue the preparation
process.
50 Error
Frequency inverter fault!
Fault code: 0.0
An error in the frequency
inverter is detected.
Re-start the machine.
If the error remains, contact
Struers Service.
Please make a note of the
Fault code.
53 Error
Specimen mover motor
power supply out of range
or missing!
Re-start the machine.
If the error remains, contact
Struers Service.
55 Error
No communication to
frequency inverter!
Re-start the machine.
If the error remains, contact
Struers Service.
58 Warning
A bad electrical connection
for the following output is
detected:
Re-start the machine.
Make a note of the output.
If the error remains, contact
Struers Service.
59 Warning
Specimen holder motor
overload, please reduce the
force.
Reduce the force or/and
increase the specimen
rpms.
Re-start the process.
If the error remains, contact
Struers Service.
8 Troubleshooting
LaboForce-100 27