28
Steris System
Caution Always install the soaking cap prior to sterilization.
1. Clean and prepare the camera head and cable as recommended
in the Cleaning and Disinfection sections. Ensure the soaking cap
is installed.
2. Sterilize the camera head and cable using Steris® System 1 with
Steris® Sterilant 20.
3. Allow the camera head, cable, coupler, and scope to completely
dry before reassembly.
Note Be sure the threads on the coupler are completely dry
before reattaching the coupler to the camera head. Any
moisture on the threads will cause the camera and coupler
windows to fog during use.
Sterrad System
Caution Always install the soaking cap prior to sterilization.
1. Clean and prepare the camera head and cable as recommended
in the Cleaning and Disinfection sections. Ensure the soaking cap
is installed.
2. Sterilize the camera head and cable using the Sterrad™
Sterilization System.
3. Allow the camera head, cable, coupler, and scope to completely
dry before reassembly.
Note Be sure the threads on the coupler are completely dry
before reattaching the coupler to the camera head. Any
moisture on the threads will cause the camera and coupler
windows to fog during use.