30
Problem Possible Solution
No color bar Ensure the video-out from the camera is
connected to the video-in on the monitor.
Ensure all video systems are powered on.
Incorrect picture
color
Perform the white balance procedure. (See the
“Using the White Balance Switch” section of this
manual.)
Check the color settings on the monitor.
Picture is too
dark
Adjust the light level with the camera head.
Increase the light-source output.
Check the fiberoptic light cable for excessive
broken fibers.
Press the GAIN switch to increase illumination.
Picture is too
bright
Adjust the light level with the camera head.
Decrease the light-source output.
Ensure GAIN is OFF.
Noise or snow on
picture when
using
electrocautery
probes
Plug the electrocautery generator into a
separate electrical outlet and separate the 1088
power cord from the electrocautery power cord.
Separate the camera cable from the
electrocautery cable
Reposition the electrocautery grounding pad on
the patient.
Noise or snow on
picture when not
using
electrocautery
probes
Ensure GAIN is OFF.
Reduce Enhancement.
Check for and replace faulty video cables.
No video picture
when the camera
head is plugged
in
Check to ensure that all devices in the video
system are plugged in and powered on.
Check the connector on the camera-head cable
for broken pins.
Troubleshooting