Service Information
20
OXYGEN BOTTLE HOLDER REPLACEMENT
Required Tools:
3/8” Drive Ratchet 1/2” Socket
1. Using the foot pedal, raise the litter to maximum height. Raise the Fowler to 90 degrees.
2. Remove the 3 bolts (A) holding the O2 bottle bracket (B) to the litter frame on the patient’s left side and
discard the bolts and bottle holder assembly.
3. Use the 3 bolts provided in the kit to attach the new bottle holder assembly to the litter frame.