EasyManua.ls Logo

Stryker 3000 - 1.7 Preventative Maintenance

Stryker 3000
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Weigh System Usage
16
OPERATING THE SCALE BEFORE PUTTING A NEW PATIENT IN BED
S Prepare bed for patient stay (linens, pillows, etc.).
S Press and release SCALE ON. The scale monitor will read:
WEIGHING
XXX.X LB
S Press and hold ZERO. The scale monitor will read:
HOLD TO ZERO WT.
RELEASE TO ZERO
S Release ZERO. The scale monitor will now read:
DO NOT TOUCH BED
0.0 LB
The bed is now ready for the patient.
NOTE
Do not zero the bed while a patient is in bed. If this should occur, remove the patient and zero the bed.
OPERATING THE SCALE IF PATIENT IS ALREADY IN BED
S If it is necessary to add or remove special equipment (monitors, pumps, etc.) during the patients
stay, press and release SCALE ON to activate the weigh system. After the scale monitor reads
XXX.X LB, press and hold . The scale monitor will read:
HOLD TO START
RELEASE TO START
S Release . The scale monitor will read:
DO NOT TOUCH BED
ADD/REMOVE EQUIP
S Add or remove the equipment and press . The scale monitor will read:
RELEASE TO FIN.
S Release . The scale monitor will read:
DO NOT TOUCH BED
XXX.X LB
The weight displayed will be that of the patient only.

Other manuals for Stryker 3000

Related product manuals