8 www.stryker.com
EN 0210-100-700 Rev-T
Specifications (continued)
Guidance and manufacturer’s declaration - electromagnetic immunity
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment specified below.
The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series should assure that it is used in such an
environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines ±2 kV on DC power
lines
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV for input/output lines
Surge
IEC 61000-4-5
±1 kV differential mode
Not Applicable Not Applicable
±2 kV common mode
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
Voltage dips, short interruptions and
voltage variations on power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
Not Applicable Not Applicable
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
NOTE: U
T
is the alternating current mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications equipment and the
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=1.2√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to 2.5 GHz
d=2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1. 2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
0000004059, Rev. T Effective Date: Sep 26, 2016 2:47:57 PM
Print Date: Nov 02, 2016 02:57:27 PM