1-10 2971-209-001 REV C www.stryker.com
Installation
The product must reach the ambient operating temperature before installation or operation to avoid the risk of
permanent damage to this product. Upon initial installation, reset the CPR.
WARNING
Risk of Shock. This product must only be connected to a supply mains with protective earth. Grounding reliability
can be achieved only when a hospital grade receptacle is used.
Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal, mask type, or half bed-length tent type.
Do not move the support surface without assistance.
Do not operate the support surface without the Isolibrium top and bottom covers and linens.
Always position the support surface on the InTouch bed frame to that the support surface does not contact the
headboard.
Do not use when gaps between the support surface and the bed frame are present to prevent the risk of
entrapment. Only use the Isolibrium support surface with InTouch bed frames.
Do not use fitted sheets with this support surface.
Do not attach the power cord to any moving parts of the InTouch bed frame. Improper routing of the power cord
could cause the risk of electric shock, pinching the power cord, or interfere with bed functions.
Always confirm proper functionality of the scale system after you install the support surface to prevent product
malfunction.
Do not use extension cords with Isolibrium. It is only intended to be powered by InTouch with the power cord
supplied.
Always use the power source provided by the InTouch foot end auxiliary power outlet with Isolibrium.
Note: Do not plug into the power source until the assembly is complete.
Prerequisite: Minimum of two operators are required to install the support surface.
1. Put the InTouch bed into the flat position before installing the support surface.
2. If powered, unplug the InTouch from the wall outlet and turn the battery switch to off.
3. Remove the InTouch headboard and footboard from the bed.
4. Position an operator at each end of the support surface and lift onto the InTouch bed frame.

Table of Contents

Other manuals for Stryker Isolibrium 2971

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Isolibrium 2971 and is the answer not in the manual?

Stryker Isolibrium 2971 Specifications

General IconGeneral
BrandStryker
ModelIsolibrium 2971
CategoryMedical Equipment
LanguageEnglish

Related product manuals