Changing Setup Options
LIFEPAK 1000 Defibrillator Operating Instructions D-1
Copyright ©2021 Stryker
CHANGING SETUP OPTIONS
Setup options allow you to define operating features for your defibrillator, such as CPR intervals.
Setup options are listed in tables beginning with Tab le D- 1 .
To enter Setup mode:
1 Ensure that no electrodes or cables are connected to the defibrillator.
2 Press and hold both softkeys and press the
ON/OFF button. The Enter Setup Mode screen
appears.
Figure D-1 Enter Setup Mode
3 Enter the Setup mode passcode. The factory default passcode is 0000—press the MENU
button four times to accept the default passcode. For information on how to change the
passcode, see page D-6.
Note: To exit Setup mode, turn the defibrillator off. If you changed the setup options, the
changes are saved and will appear the next time you turn the defibrillator on. (Refer to Setup
menu options that follow.)
Setup
Increase
Decrease
Press button to set
Enter Setup Mode passcode