LIFEPAK
®
1000 DEFIBRILLATOR
USER’S CHECKLIST
Unit Serial Number __________________________
Department/Location________________________
Instruction
Recommended
Corrective Action
Date
Initials
1 Check readiness
display for:
WRENCH symbol Contact authorized service
personnel.
OK symbol None.
Battery level Replace if one bar or less on
battery status indicator.
2 Check Use By date on
electrode packet.
Replace electrode packet if date
has passed.
3 Check spare battery. Ensure spare, fully charged
battery is available.
4 Check additional
supplies.
Replenish as needed.
5 Check defibrillator
for:
Damage or cracks Contact authorized service
personnel.
Foreign substances Clean the device.
6 Comments: