Electromagnetic Compatibility Guidance
LIFEPAK 1000 Defibrillator Operating Instructions G-5
Copyright ©2021 Stryker
1700-2100
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3, 4, 25
UMTS
DCS-1800
PCS-1900
UTRA Band 1, 2, 4, 9
2.0 0.3
d = 6/9
2400-2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
4G/LTE-A
2.0 0.3
d = 6/28
5100-5800 WLAN 802.11 a/n 0.2 0.3
d = 6/9
For transmitters in the frequency ranges listed above, the recommended separation distance d in
meters (m) can be determined using the equation in the table above with the known power, P of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Band (MHz) Radio Service
Maximum
Power (W)
Distance (m) Equation