LIFEPAK 15 Monitor/Debrillator
EtCO
2
monitoring procedure
1. Press ON
2. Select appropriate EtCO
2
accessory for the patient
3. Open CO
2
port door and insert FilterLine connector,
turn connector clockwise until tight
4. Verify CO
2
area is displayed
5. Display CO
2
waveform in Channel 2 or 3
6. Connect FilterLine set to the patient
7. Conrm the EtCO
2
waveform is displayed
Auto-zero is part of initialization self-test performed
by monitor. Scale is automatically selected for
best visualization.
Note: If you use a ventilation system, do not connect FilterLine set
to the patient/ventilation system until self-test is complete
CO2