LIFEPAK 15 Monitor/Debrillator
SpO
2
1. Press ON
2. Connect the pulse
oximeter cable to the
monitor and sensor
3. Attach sensor to ring
nger of the non-
dominant hand and
orient so cable is on
back of patient’s hand
4. Observe the pulse bar
for uctuation, amplitude
of the pulse bar indicates
relative signal quality
5. Conrm SpO
2
reading
appears and is stable
SpCO & SpMet
Perform steps 1-5, then:
6. Verify that SpCO / SpMet
sensor is in use; ONLY
Rainbow
®
sensors are
capable of reading
SpCO / SpMet
7. Rotate Speed Dial to
outline SpO
2
area
8. Select parameter
9. Select SpCO or SpMet.
Selected value displays
for 10 seconds.
CO2
SpO2
NIBP
ECG
P1
P2
SpO
2
/ SpCO / SpMet
monitoring procedure
Note: If SpCO or SpMet reading is above
normal limits (SpCO >10% , SpMet >3%)
an Advisory message occurs.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LIFEPAK 15 and is the answer not in the manual?

Stryker LIFEPAK 15 Specifications

General IconGeneral
BatteryLithium-ion
DefibrillatorYes, biphasic
Defibrillation Energy Range1-360 J
SpO2 MonitoringYes
NIBP MonitoringYes
EtCO2 MonitoringYes
Invasive Pressure MonitoringYes
Temperature MonitoringYes
ConnectivityBluetooth, WiFi
ECG Monitoring12-lead