Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15 Monitor/Debrillator
EtCO
2
continued
Observation Action
NO BREATH ALARM
message appears and
waveform is solid line
at or near zero
Check the patient
Twist connecter clockwise
until snug
Check for leaks or disconnected
tubing
There is no EtCO
2
value
and the waveform is at
Measured CO
2
is <3.5mmHg. See
detection above.
CO
2
FilterLine OFF
message appears and
waveform is (---)
Connect FilterLine set to device port
Twist FilterLine connector
clockwise until snug and
rmly seated
CO
2
FilterLine PURGING
message appears and
waveform is (---)
Disconnect then reconnect the
FilterLine set
Twist FilterLine connector
clockwise until snug and
rmly seated
CO
2
FilterLine
BLOCKAGE message
appears and waveform
is (---)
Disconnect, reconnect
If unsuccessful, change the
FilterLine set
XXX appears instead of
EtCO
2
value
Turn device off then on again. If
problem persists, contact qualied
service personnel.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LIFEPAK 15 and is the answer not in the manual?

Stryker LIFEPAK 15 Specifications

General IconGeneral
BatteryLithium-ion
DefibrillatorYes, biphasic
Defibrillation Energy Range1-360 J
SpO2 MonitoringYes
NIBP MonitoringYes
EtCO2 MonitoringYes
Invasive Pressure MonitoringYes
Temperature MonitoringYes
ConnectivityBluetooth, WiFi
ECG Monitoring12-lead

Related product manuals