Page 26
Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15 Monitor/Debrillator
Invasive Pressure (IP)
Observation Action
Invasive pressure
value is blank
Connect the transducer to the cable and the
cable to the monitor
PX NOT ZEROED
message appears
Zero the transducer
PX ZERO FAILED
message appears
Conrm the transducer is open to air and
repeat attempt to zero
Dampened
waveform
Check the entire system for leaks. Tighten all
connections. Replace any defective stopcocks.
Use syringe to draw back air or particles in
catheter, then ush system
No waveform or
pressure reading
Check patient. Check stopcock positions and
monitor setup.
Invasive BP lower
than cuff BP
Reposition transducer to correct height
Tighten all connections
Flush system
Open stopcock to air and rezero transducer
Invasive BP higher
than cuff BP
Reposition transducer to correct height
Rezero
Change catheter tip position
Use mean pressure values
Inability to zero
system
Check the stopcock position. Replace any
defective stopcocks.
Replace transducer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LIFEPAK 15 and is the answer not in the manual?

Stryker LIFEPAK 15 Specifications

General IconGeneral
BatteryLithium-ion
DefibrillatorYes, biphasic
Defibrillation Energy Range1-360 J
SpO2 MonitoringYes
NIBP MonitoringYes
EtCO2 MonitoringYes
Invasive Pressure MonitoringYes
Temperature MonitoringYes
ConnectivityBluetooth, WiFi
ECG Monitoring12-lead

Related product manuals