Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15 Monitor/Debrillator
Temperature
Observation Action
Screen remains blank Temp outside range of 24.8° and
45.2° C (76.6° and 113.4° F)
CHECK SENSOR message
appears and value is (---)
Temp outside of range, the probe
is dislodged, not connected or the
cable not connected to the monitor
Check that probe is properly
positioned
CHECK SENSOR message
appears while value is
displayed
Temp probe is dislodged and value
is below 31° C (87.8° F) or above
41.0° C (105.8° F)
TEMP: ACCURACY
OUTSIDE LIMITS message
appears and value is XXX
Temperature module is not
calibrated or accuracy check failed
Turn device off then on again. If
problem persists, contact qualied
service personnel.
NOTE: The temperature probe may require 3 minutes
to equilibrate after placement on the patient
monitoring site.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker LIFEPAK 15 and is the answer not in the manual?

Stryker LIFEPAK 15 Specifications

General IconGeneral
BatteryLithium-ion
DefibrillatorYes, biphasic
Defibrillation Energy Range1-360 J
SpO2 MonitoringYes
NIBP MonitoringYes
EtCO2 MonitoringYes
Invasive Pressure MonitoringYes
Temperature MonitoringYes
ConnectivityBluetooth, WiFi
ECG Monitoring12-lead

Related product manuals