Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The VisionPro SYNK 26” Wireless LED Display and SYNK Wireless Transmitter are intended for use inthe electromagnetic
environment specied below� The customer or the user of the VisionPro SYNK 26” Wireless LED Display and SYNK Wireless
Transmitter should ensure they are used in such an environment�
Emissions test Compliance Electromagnetic Environment - guidance
RF emissions CISPR 11 Group 1
The VisionPro SYNK 26” Wireless LED Display and SYNK
Wireless Transmitter uses RF energy only for its internal
function; therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment�
RF emissions CISPR 11 Class B The VisionPro SYNK 26” Wireless LED Display and
SYNK Wireless Transmitter is suitable for use in all
establishments other than domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes, provided the following warning is
heeded:
Warning: This system is intended for use by
health care professionals only. This system may
cause radio interference or may disrupt the op-
eration of nearby equipment. It may be necessary
to take mitigation measures, such as reorienting
or relocating the system or shielding the location.
Harmonic emissions IEC61000-3-2 Class D
Voltage Fluctuations/ icker
emissions IEC61000-3-3
Complies
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
The VisionPro SYNK 26” Wireless LED Display and SYNK Wireless Transmitter are intended for use in the electromagnetic
environment specied below� The customer or the user of the VisionPro SYNK 26” Wireless LED Display and SYNK Wireless
Transmitter should ensure that they are used in such an environment�
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment--
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floors should be wood, concrete,
or ceramic tile� If oors are covered
with synthetic material, the relative
humidity should be at least 30%�
Electrical fast
transient/burst
IEC61000-4-4
2kV for power supply lines
1kV for input/output lines
(if applicable)
2kV line to ground
1kV line to line
(if applicable)
Mains power quality should be that
of a typical commercial or hospital
environment�
Surge
IEC61000-4-5
1kV dierential mode
2kV common mode
1kV dierential mode
2kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
0% U
T
0�5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 sec�
0% U
T
0�5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 sec�
Mains power quality should be
that of a typical commercial or
hospital environment� If the user of
the transmitter requires continued
operation during power mains
interruptions, it is recommended that
the Wireless Transmitter be powered
from an uninterruptible power supply
or a battery�
Power frequency
(50/60Hz) magnetic
eld
IEC 61000-4-8
30 A/m 30 A/m Power-frequency magnetic elds
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment�
NOTE: U
T
is the AC mains voltage prior to application of the test level�